NP330-F Water in oil sensor
Measure water in oil
Whether in hydraulic oils, diesel, vegetable oils, Skydrol or other user-specific fluids: the NP330-F measures the water content up to the saturation limit. Specially calibrated for your application and the oils and fluids used – for example in drying and oil treatment plants, production plants for vegetable oils, hydraulic systems and especially for monitoring oil coolers.
NP330-F
How we measure success: Your advantages.
Custom calibration
Calibration on the user-specific fluids and requirements
Temperature compensation
Elimination of influence of temperature changes on the measured value
High measuring accuracy
Through multi-point calibration and temperature compensation
Fast reaction time
Fast response to humidity change
High compressive strength
Patented pressure resistance up to 300 bar
Long product life cycles
Designed for industrial use and long-term spare parts availability
Technology at a glance
See here how the water in oil sensor is constructed and which components make up its functionality. We will be happy to show you more, live or online, during a consultation.
Technical information
Inline measurement of the absolute water content
The NP330-F water in oil sensor is designed for in-line measurement of absolute water content in ppm [mg/kg] below the saturation limit. The sensor performs its work in mineral, ester, skydrol, silicone, vegetable, diesel, refrigerator, transformer and insulating oils, among others, as well as in other customer-specific fluids.
The fields of application
- Drying and oil treatment plants
- Production plants for vegetable oils
- Painting lines for automotive parts
- Oil cooler monitoring
- Hydraulic systems
- Refrigerator oil
- Hardening shops
- Aircraft industry
Current output | 4–20 mA |
Serial interface | RS232 |
Switching output | 24 V DC |
Typical measuring range | 10 – 40,000 ppm (upper range limited to saturation of the oil) |
Temperature liquid | -20°C to +70°C (higher media temperatures on request) |
Supply | 18 – 24 V DC / max. 70 mA |
System pressure | up to 300 bar |
Installation depth | 50 mm with ½-inch thread |
Protection class | IP65 |
Service
How is the product maintained?
We recommend an annual recalibration and adjustment for all our sensors to be able to guarantee the exact functionality. We calibrate our sensors with state-of-the-art measurement technology in our in-house calibration laboratory. Our test equipment is one of the most accurate references in measurement and control technology. The test results are summarised in a factory calibration certificate in accordance with DIN EN ISO/IEC 17025:2018-03.